Recall Z-2296-2020— BioFire Diagnostics, LLC

Class II · Terminated

Class II FDA medical device recall by BioFire Diagnostics, LLC, initiated 2020-01-22, terminated 2021-04-14. It names 4 FDA 510(k) clearances: K181493, K160457, K143171 and 1 more. Product: FilmArray BCID Panel, RFIT-ASY-0127, 6, Rx Only, CE, IVD, UDI: 00815381020093 used with the following BACT/ALERT Blood Culture Bottle Catalog No./Description: 410851/BACT/ALERT¿ FA Plus 410852/BACT/ALERT¿ FN Plus 410853/BACT/ALERT¿ PF Plus 259790/BACT/ALERT¿ SN Product Usage: is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.. Reason: There is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive Proteus results from this test may lead to patients receiving improper antibiotic treatment until culture results are available (within 24-48 hours)..

Recalling firm
BioFire Diagnostics, LLC
Classification
Class II
Status
Terminated
Initiated
2020-01-22
Terminated
2021-04-14
Firm location
Salt Lake City, UT, United States

510(k) clearances named in this recall

4 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FilmArray BCID Panel, RFIT-ASY-0127, 6, Rx Only, CE, IVD, UDI: 00815381020093 used with the following BACT/ALERT Blood Culture Bottle Catalog No./Description: 410851/BACT/ALERT¿ FA Plus 410852/BACT/ALERT¿ FN Plus 410853/BACT/ALERT¿ PF Plus 259790/BACT/ALERT¿ SN Product Usage: is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.

Reason for recall

There is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive Proteus results from this test may lead to patients receiving improper antibiotic treatment until culture results are available (within 24-48 hours).