Recall Z-2294-2025— Merit Medical Systems, Inc.
Class II · Ongoing
Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2025-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029. Reason: High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient..
- Recalling firm
- Merit Medical Systems, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-06-05
- Firm location
- South Jordan, UT, United States
Product description
Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
Reason for recall
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.