Recall Z-2292-2016— Zimmer Gmbh
Class II · Terminated
Class II FDA medical device recall by Zimmer Gmbh, initiated 2016-06-30, terminated 2017-03-27. It names one FDA 510(k) clearance: K123392. Product: Avenir M¿ller Stem 4 Standard Avenir M¿ller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended Use: " Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases. " Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR). " Acute traumatic fracture of the femoral head or neck " Avascular necrosis of the femoral head. " Avenir Moller Stems are for cementless use only.. Reason: Product mix-up. The Avenir M¿ller Stem 6 lateral uncemented might be placed in the packaging of the Avenir M¿ller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of incorrect implant which may in turn lead dismetry in patient and possibly the need for revision..
- Recalling firm
- Zimmer Gmbh
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-06-30
- Terminated
- 2017-03-27
- Firm location
- Winterthur, N/A, Switzerland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Avenir M¿ller Stem 4 Standard Avenir M¿ller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended Use: " Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases. " Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR). " Acute traumatic fracture of the femoral head or neck " Avascular necrosis of the femoral head. " Avenir Moller Stems are for cementless use only.
Reason for recall
Product mix-up. The Avenir M¿ller Stem 6 lateral uncemented might be placed in the packaging of the Avenir M¿ller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of incorrect implant which may in turn lead dismetry in patient and possibly the need for revision.