Recall Z-2290-2020— Qiagen Sciences LLC

Class II · Terminated

Class II FDA medical device recall by Qiagen Sciences LLC, initiated 2020-04-08, terminated 2021-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121. Reason: There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result..

Recalling firm
Qiagen Sciences LLC
Classification
Class II
Status
Terminated
Initiated
2020-04-08
Terminated
2021-06-03
Firm location
Germantown, MD, United States

Product description

therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121

Reason for recall

There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.