Recall Z-2289-2015— The Anspach Effort, Inc.

Class II · Terminated

Class II FDA medical device recall by The Anspach Effort, Inc., initiated 2015-04-13, terminated 2016-08-25. It names one FDA 510(k) clearance: K113476. Product: ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.. Reason: G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater..

Recalling firm
The Anspach Effort, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-04-13
Terminated
2016-08-25
Firm location
Palm Beach Gardens, FL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.

Reason for recall

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.