Recall Z-2286-2018— Horiba Instruments, Inc dba Horiba Medical

Class II · Terminated

Class II FDA medical device recall by Horiba Instruments, Inc dba Horiba Medical, initiated 2018-04-26, terminated 2019-09-26. FDA's recall record names no specific 510(k) clearance for this event. Product: ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.. Reason: When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types..

Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Classification
Class II
Status
Terminated
Initiated
2018-04-26
Terminated
2019-09-26
Firm location
Irvine, CA, United States

Product description

ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Reason for recall

When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.