Recall Z-2279-2013— CareFusion 303, Inc.

Class II · Terminated

Class II FDA medical device recall by CareFusion 303, Inc., initiated 2013-08-30, terminated 2013-10-28. It names one FDA 510(k) clearance: K931550. Product: SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors.. Reason: CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead..

Recalling firm
CareFusion 303, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-08-30
Terminated
2013-10-28
Firm location
San Diego, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors.

Reason for recall

CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.