Recall Z-2276-2024— American Contract Systems Inc

Class II · Ongoing

Class II FDA medical device recall by American Contract Systems Inc, initiated 2024-05-15. FDA's recall record names no specific 510(k) clearance for this event. Product: MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B. Reason: Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012..

Recalling firm
American Contract Systems Inc
Classification
Class II
Status
Ongoing
Initiated
2024-05-15
Firm location
Tiffin, OH, United States

Product description

MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B

Reason for recall

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Data sourced from openFDA. This site is unofficial and independent of the FDA.