Recall Z-2274-2016— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2016-02-05, terminated 2017-01-17. FDA's recall record names no specific 510(k) clearance for this event. Product: Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.. Reason: A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2016-02-05
Terminated
2017-01-17
Firm location
Brea, CA, United States

Product description

Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.

Reason for recall

A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.

Data sourced from openFDA. This site is unofficial and independent of the FDA.