Recall Z-2273-2023— Maquet Cardiovascular, LLC
Class II · Ongoing
Class II FDA medical device recall by Maquet Cardiovascular, LLC, initiated 2023-06-28. FDA's recall record names no specific 510(k) clearance for this event. Product: ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.. Reason: The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion..
- Recalling firm
- Maquet Cardiovascular, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-06-28
- Firm location
- Wayne, NJ, United States
Product description
ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.
Reason for recall
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Data sourced from openFDA. This site is unofficial and independent of the FDA.