Recall Z-2267-2019— Degania Silicone, Ltd.

Class II · Terminated

Class II FDA medical device recall by Degania Silicone, Ltd., initiated 2019-06-28, terminated 2020-10-22. FDA's recall record names no specific 510(k) clearance for this event. Product: Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.. Reason: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body.

Recalling firm
Degania Silicone, Ltd.
Classification
Class II
Status
Terminated
Initiated
2019-06-28
Terminated
2020-10-22
Firm location
Dganya Bet, N/A, Israel

Product description

Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Reason for recall

15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

Data sourced from openFDA. This site is unofficial and independent of the FDA.