Recall Z-2266-2024— Ortho-Clinical Diagnostics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2024-05-21. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.. Reason: VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides..
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-05-21
- Firm location
- Rochester, NY, United States
Product description
VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.
Reason for recall
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.