Recall Z-2265-2018— TriMed Inc.

Class II · Terminated

Class II FDA medical device recall by TriMed Inc., initiated 2018-04-13, terminated 2021-01-22. It names one FDA 510(k) clearance: K112794. Product: Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15. Reason: During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized..

Recalling firm
TriMed Inc.
Classification
Class II
Status
Terminated
Initiated
2018-04-13
Terminated
2021-01-22
Firm location
Santa Clarita, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15

Reason for recall

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Data sourced from openFDA. This site is unofficial and independent of the FDA.