Recall Z-2263-2012— Cisco Systems Inc
Class II · Terminated
Class II FDA medical device recall by Cisco Systems Inc, initiated 2012-07-23, terminated 2012-08-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party EMR: scheduling: and other software applications for remote consultations and exams. Cisco Health Presence is not intended for use in emergency situations. Cisco HealthPresence is not intended for use in active patient monitoring or alarming.. Reason: Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs readings will only update on the Cisco HealthPresence Vitals tab and within the OnePlace database, but will not update under the OnePlace tab, and thus, will not provide current vit.
- Recalling firm
- Cisco Systems Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-07-23
- Terminated
- 2012-08-24
- Firm location
- San Jose, CA, United States
Product description
Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party EMR: scheduling: and other software applications for remote consultations and exams. Cisco Health Presence is not intended for use in emergency situations. Cisco HealthPresence is not intended for use in active patient monitoring or alarming.
Reason for recall
Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs readings will only update on the Cisco HealthPresence Vitals tab and within the OnePlace database, but will not update under the OnePlace tab, and thus, will not provide current vit
Data sourced from openFDA. This site is unofficial and independent of the FDA.