Recall Z-2262-2016— The Metrix Company
Class II · Terminated
Class II FDA medical device recall by The Metrix Company, initiated 2016-06-23, terminated 2023-04-11. It names one FDA 510(k) clearance: K926123. Product: Dual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mL. Reason: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications..
- Recalling firm
- The Metrix Company
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-06-23
- Terminated
- 2023-04-11
- Firm location
- Dubuque, IA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mL
Reason for recall
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.