Recall Z-2260-2015— Medtronic MiniMed Inc.

Class II · Terminated

Class II FDA medical device recall by Medtronic MiniMed Inc., initiated 2015-06-19, terminated 2015-10-02. FDA's recall record names no specific 510(k) clearance for this event. Product: MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.. Reason: Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate..

Recalling firm
Medtronic MiniMed Inc.
Classification
Class II
Status
Terminated
Initiated
2015-06-19
Terminated
2015-10-02
Firm location
Northridge, CA, United States

Product description

MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.

Reason for recall

Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate.

Data sourced from openFDA. This site is unofficial and independent of the FDA.