Recall Z-2258-2024— GE Medical Systems Information Technologies Inc

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems Information Technologies Inc, initiated 2024-04-22. It names one FDA 510(k) clearance: K050093. Product: Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).. Reason: Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms..

Recalling firm
GE Medical Systems Information Technologies Inc
Classification
Class II
Status
Ongoing
Initiated
2024-04-22
Firm location
Milwaukee, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).

Reason for recall

Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

Data sourced from openFDA. This site is unofficial and independent of the FDA.