Recall Z-2258-2016— The Metrix Company

Class II · Terminated

Class II FDA medical device recall by The Metrix Company, initiated 2016-06-23, terminated 2023-04-11. FDA's recall record names no specific 510(k) clearance for this event. Product: AUTOCOMP6 XP High Speed Compounder REF 58810. Reason: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications..

Recalling firm
The Metrix Company
Classification
Class II
Status
Terminated
Initiated
2016-06-23
Terminated
2023-04-11
Firm location
Dubuque, IA, United States

Product description

AUTOCOMP6 XP High Speed Compounder REF 58810

Reason for recall

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Data sourced from openFDA. This site is unofficial and independent of the FDA.