Recall Z-2257-2024— Nalu Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Nalu Medical, Inc., initiated 2024-06-03. It names 2 FDA 510(k) clearances: K183579, K183047. Product: Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.. Reason: Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain..
- Recalling firm
- Nalu Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-06-03
- Firm location
- Carlsbad, CA, United States
510(k) clearances named in this recall
2 on FDA's record- K183579 — Nalu Neurostimulation Kit (Integrated, 40 cm: Single 8/Dual 8), Nalu Neurostimulation Kit (Ported, 2 cm: Single 8/Dual 8), Dual 8 Ported Nalu Implantable Pulse Generator with 40 cm Kit, 40 cm/ 60 cm Trial/Extension Lead Kits, Patient Kits and miscellaneous replacement kits
- K183047 — Nalu Neurostimulation System
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.
Reason for recall
Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.