Recall Z-2256-2012— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2012-07-10, terminated 2013-01-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body.. Reason: Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2012-07-10
Terminated
2013-01-10
Firm location
Waukesha, WI, United States

Product description

Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body.

Reason for recall

Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.