Recall Z-2255-2018— Cyberonics, Inc
Class II · Terminated
Class II FDA medical device recall by Cyberonics, Inc, initiated 2017-11-13, terminated 2022-10-31. FDA's recall record names no specific 510(k) clearance for this event. Product: VNS Therapy Programmer, Model 3000, v1.0 System. Reason: Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106)..
- Recalling firm
- Cyberonics, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-11-13
- Terminated
- 2022-10-31
- Firm location
- Houston, TX, United States
Product description
VNS Therapy Programmer, Model 3000, v1.0 System
Reason for recall
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
Data sourced from openFDA. This site is unofficial and independent of the FDA.