Recall Z-2248-2017— Stryker Communications

Class II · Terminated

Class II FDA medical device recall by Stryker Communications, initiated 2017-04-21, terminated 2017-09-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated. Reason: The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury..

Recalling firm
Stryker Communications
Classification
Class II
Status
Terminated
Initiated
2017-04-21
Terminated
2017-09-20
Firm location
Flower Mound, TX, United States

Product description

Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated

Reason for recall

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.