Recall Z-2247-2018— Synthes (USA) Products LLC

Class II · Terminated

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2018-03-05, terminated 2020-02-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Part no. 690.380, Graphic Case F/7.3mm Cannulated Screw Set. Reason: Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Terminated
Initiated
2018-03-05
Terminated
2020-02-28
Firm location
West Chester, PA, United States

Product description

Part no. 690.380, Graphic Case F/7.3mm Cannulated Screw Set

Reason for recall

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Data sourced from openFDA. This site is unofficial and independent of the FDA.