Recall Z-2245-2024— Ortho-Clinical Diagnostics, Inc.

Class III · Ongoing

Class III FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2024-05-14. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344. Reason: Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately..

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Classification
Class III
Status
Ongoing
Initiated
2024-05-14
Firm location
Rochester, NY, United States

Product description

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344

Reason for recall

Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.

Data sourced from openFDA. This site is unofficial and independent of the FDA.