Recall Z-2245-2024— Ortho-Clinical Diagnostics, Inc.
Class III · Ongoing
Class III FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2024-05-14. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344. Reason: Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately..
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Classification
- Class III
- Status
- Ongoing
- Initiated
- 2024-05-14
- Firm location
- Rochester, NY, United States
Product description
VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344
Reason for recall
Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.
Data sourced from openFDA. This site is unofficial and independent of the FDA.