Recall Z-2242-2020— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2019-04-30, terminated 2021-12-28. It names one FDA 510(k) clearance: K112933. Product: Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.. Reason: The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2019-04-30
Terminated
2021-12-28
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Reason for recall

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Data sourced from openFDA. This site is unofficial and independent of the FDA.