Recall Z-2240-2020— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2019-04-30, terminated 2021-12-28. It names 2 FDA 510(k) clearances: K031269, K002905. Product: Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.. Reason: The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-04-30
- Terminated
- 2021-12-28
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
2 on FDA's record- K031269 — ACCESS THYROGLOBULIN REAGENTS ON THE ACCESS IMMUNOASSAY SYSTEMS
- K002905 — ACCESS THYROGLOBULIN REAGENTS ON THE ACCESS ANALYZER
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
Reason for recall
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.