Recall Z-2238-2016— Stryker Instruments Div. of Stryker Corporation

Class II · Terminated

Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2016-06-24, terminated 2017-03-23. It names one FDA 510(k) clearance: K143399. Product: 2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.. Reason: A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur..

Recalling firm
Stryker Instruments Div. of Stryker Corporation
Classification
Class II
Status
Terminated
Initiated
2016-06-24
Terminated
2017-03-23
Firm location
Portage, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.

Reason for recall

A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.