Recall Z-2234-2023— Boston Scientific Corporation

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2023-05-25. It names one FDA 510(k) clearance: K210925. Product: Flexiva Pulse ID 242 TracTip single Use Fiber-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy 5-pack UPN: M006L8406961. Reason: Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched.

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Ongoing
Initiated
2023-05-25
Firm location
Marlborough, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Flexiva Pulse ID 242 TracTip single Use Fiber-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy 5-pack UPN: M006L8406961

Reason for recall

Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched

Data sourced from openFDA. This site is unofficial and independent of the FDA.