Recall Z-2230-2025— Siemens Healthcare Diagnostics Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Healthcare Diagnostics Inc, initiated 2025-06-26. It names one FDA 510(k) clearance: K002738. Product: RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.. Reason: Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients..
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-06-26
- Firm location
- Norwood, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Reason for recall
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.