Recall Z-2228-2026— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2026-04-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039), Cardio Vascular-Allura (722400);. Reason: The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2026-04-28
Firm location
Best, Netherlands

Product description

Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039), Cardio Vascular-Allura (722400);

Reason for recall

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Data sourced from openFDA. This site is unofficial and independent of the FDA.