Recall Z-2228-2023— Boston Scientific Corporation
Class II · Ongoing
Class II FDA medical device recall by Boston Scientific Corporation, initiated 2023-05-25. It names one FDA 510(k) clearance: K210925. Product: Flexiva Pulse 242 Single-Use Laser Fibers-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Single-pack UPN : M006L8405910. Reason: Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched.
- Recalling firm
- Boston Scientific Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-05-25
- Firm location
- Marlborough, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Flexiva Pulse 242 Single-Use Laser Fibers-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Single-pack UPN : M006L8405910
Reason for recall
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Data sourced from openFDA. This site is unofficial and independent of the FDA.