Recall Z-2224-2026— Fresenius Kabi USA, LLC

Class I · Ongoing

Class I FDA medical device recall by Fresenius Kabi USA, LLC, initiated 2026-05-06. It names one FDA 510(k) clearance: K183311. Product: Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004. Reason: Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern..

Recalling firm
Fresenius Kabi USA, LLC
Classification
Class I
Status
Ongoing
Initiated
2026-05-06
Firm location
North Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Reason for recall

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Data sourced from openFDA. This site is unofficial and independent of the FDA.