Recall Z-2223-2017— Roche Diagnostics Corporation

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2017-05-09, terminated 2018-02-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system. Reason: a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks..

Recalling firm
Roche Diagnostics Corporation
Classification
Class II
Status
Terminated
Initiated
2017-05-09
Terminated
2018-02-02
Firm location
Indianapolis, IN, United States

Product description

Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system

Reason for recall

a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.