Recall Z-2220-2012— DePuy Orthopaedics, Inc.

Class II · Terminated

Class II FDA medical device recall by DePuy Orthopaedics, Inc., initiated 2012-07-24, terminated 2013-04-30. It names one FDA 510(k) clearance: K102423. Product: Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.. Reason: In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner..

Recalling firm
DePuy Orthopaedics, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-07-24
Terminated
2013-04-30
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Reason for recall

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.