Recall Z-2218-2025— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2025-06-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Upgrade to MR 7700 Model Number (REF): 782130;. Reason: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire..
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-06-30
- Firm location
- Cambridge, MA, United States
Product description
Upgrade to MR 7700 Model Number (REF): 782130;
Reason for recall
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.