Recall Z-2218-2024— Merit Medical Systems, Inc.

Class II · Ongoing

Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2024-05-22. It names one FDA 510(k) clearance: K211405. Product: Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS -PG6F45MP -PG6F45R -PG6F45S -PG6F45S018 -PG6F65MP -PG6F65S -PG6F65SX -PG6F90MP -PG6F90MPX -PG6F90SX -PG6F90SX018 -PG7F45MP -PG7F45MPX -PG7F45RX -PG7F45S -PG7F45SX -PG7F65SX -PG7F90SX -PG8F45MPX -PG8F45S -PG8F45SX -PG8F65SX -PG8F90SX -X9PG700001 The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.. Reason: Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body..

Recalling firm
Merit Medical Systems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-05-22
Firm location
South Jordan, UT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS -PG6F45MP -PG6F45R -PG6F45S -PG6F45S018 -PG6F65MP -PG6F65S -PG6F65SX -PG6F90MP -PG6F90MPX -PG6F90SX -PG6F90SX018 -PG7F45MP -PG7F45MPX -PG7F45RX -PG7F45S -PG7F45SX -PG7F65SX -PG7F90SX -PG8F45MPX -PG8F45S -PG8F45SX -PG8F65SX -PG8F90SX -X9PG700001 The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Reason for recall

Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body.