Recall Z-2216-2025— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2025-06-30. FDA's recall record names no specific 510(k) clearance for this event. Product: SmartPath to dStream for XR and 3.0T Model Numbers (REF): (1) 781270, (2) 782113, (3) 782129;. Reason: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire..

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2025-06-30
Firm location
Cambridge, MA, United States

Product description

SmartPath to dStream for XR and 3.0T Model Numbers (REF): (1) 781270, (2) 782113, (3) 782129;

Reason for recall

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.