Recall Z-2210-2013— Philips Medical Systems (Cleveland) Inc

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2013-07-05, terminated 2015-09-15. It names 2 FDA 510(k) clearances: K033326, K012009. Product: Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.. Reason: The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch..

Recalling firm
Philips Medical Systems (Cleveland) Inc
Classification
Class II
Status
Terminated
Initiated
2013-07-05
Terminated
2015-09-15
Firm location
Cleveland, OH, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.

Reason for recall

The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.

Data sourced from openFDA. This site is unofficial and independent of the FDA.