Recall Z-2209-2013— Cincinnati Sub-Zero Products Inc

Class II · Terminated

Class II FDA medical device recall by Cincinnati Sub-Zero Products Inc, initiated 2013-09-03, terminated 2014-08-22. It names one FDA 510(k) clearance: K101589. Product: Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer.. Reason: During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval..

Recalling firm
Cincinnati Sub-Zero Products Inc
Classification
Class II
Status
Terminated
Initiated
2013-09-03
Terminated
2014-08-22
Firm location
Cincinnati, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer.

Reason for recall

During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.

Data sourced from openFDA. This site is unofficial and independent of the FDA.