Recall Z-2209-2012— Roche Molecular Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Molecular Systems, Inc., initiated 2012-02-01, terminated 2013-07-16. FDA's recall record names no specific 510(k) clearance for this event. Product: COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.. Reason: Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce.

Recalling firm
Roche Molecular Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-02-01
Terminated
2013-07-16
Firm location
Branchburg, NJ, United States

Product description

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Reason for recall

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce

Data sourced from openFDA. This site is unofficial and independent of the FDA.