Recall Z-2206-2015— American Science & Engineering, Inc.
Class II · Terminated
Class II FDA medical device recall by American Science & Engineering, Inc., initiated 2015-05-22, terminated 2017-02-22. FDA's recall record names no specific 510(k) clearance for this event. Product: American Science & Engineering Inc,Gemini Cabinet X-ray system.. Reason: During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated..
- Recalling firm
- American Science & Engineering, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-05-22
- Terminated
- 2017-02-22
- Firm location
- Billerica, MA, United States
Product description
American Science & Engineering Inc,Gemini Cabinet X-ray system.
Reason for recall
During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.