Recall Z-2206-2015— American Science & Engineering, Inc.

Class II · Terminated

Class II FDA medical device recall by American Science & Engineering, Inc., initiated 2015-05-22, terminated 2017-02-22. FDA's recall record names no specific 510(k) clearance for this event. Product: American Science & Engineering Inc,Gemini Cabinet X-ray system.. Reason: During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated..

Recalling firm
American Science & Engineering, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-05-22
Terminated
2017-02-22
Firm location
Billerica, MA, United States

Product description

American Science & Engineering Inc,Gemini Cabinet X-ray system.

Reason for recall

During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.