Recall Z-2204-2019— Ad-Tech Medical Instrument Corporation

Class II · Terminated

Class II FDA medical device recall by Ad-Tech Medical Instrument Corporation, initiated 2019-06-18, terminated 2021-02-16. It names one FDA 510(k) clearance: K181544. Product: LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 13mm non-anodized anchor bolt body with anchor bolt cap containing a green gasket and silicone cap, 26mm non-anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap - Product Usage: The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.. Reason: On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts..

Recalling firm
Ad-Tech Medical Instrument Corporation
Classification
Class II
Status
Terminated
Initiated
2019-06-18
Terminated
2021-02-16
Firm location
Oak Creek, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 13mm non-anodized anchor bolt body with anchor bolt cap containing a green gasket and silicone cap, 26mm non-anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap - Product Usage: The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.

Reason for recall

On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.