Recall Z-2202-2023— MEDLINE INDUSTRIES, LP - Northfield
Class II · Ongoing
Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2023-05-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC INFANT INSERTION PACK, Model Number CVI4770A; b) INSERTION BUNDLE, Model Number DT21275; c) VA MAINTENANCE PACK, Model Number EBSI1382; d) MIDLINE DRESSING CHANGE KIT, Model Number EBSI1464;. Reason: Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams..
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-05-18
- Firm location
- Northfield, IL, United States
Product description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC INFANT INSERTION PACK, Model Number CVI4770A; b) INSERTION BUNDLE, Model Number DT21275; c) VA MAINTENANCE PACK, Model Number EBSI1382; d) MIDLINE DRESSING CHANGE KIT, Model Number EBSI1464;
Reason for recall
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.