Recall Z-2189-2016— Synthes (USA) Products LLC
Class II · Terminated
Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2016-05-31, terminated 2017-02-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short Lot # 5053859 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.. Reason: DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard..
- Recalling firm
- Synthes (USA) Products LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-05-31
- Terminated
- 2017-02-02
- Firm location
- West Chester, PA, United States
Product description
Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short Lot # 5053859 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
Reason for recall
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.