Recall Z-2188-2021— Heartware, Inc.

Class I · Ongoing

Class I FDA medical device recall by Heartware, Inc., initiated 2021-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: HeartWare HVAD Pump Implant Kit, REF 1103. Reason: Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart..

Recalling firm
Heartware, Inc.
Classification
Class I
Status
Ongoing
Initiated
2021-06-03
Firm location
Miami Lakes, FL, United States

Product description

HeartWare HVAD Pump Implant Kit, REF 1103

Reason for recall

Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.

Data sourced from openFDA. This site is unofficial and independent of the FDA.