Recall Z-2184-2021— Abbott Laboratories
Class II · Ongoing
Class II FDA medical device recall by Abbott Laboratories, initiated 2021-06-10. FDA's recall record names no specific 510(k) clearance for this event. Product: REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569. Reason: There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results..
- Recalling firm
- Abbott Laboratories
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-06-10
- Firm location
- Irving, TX, United States
Product description
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569
Reason for recall
There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.