Recall Z-2183-2024— Siemens Medical Solutions USA, Inc

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2024-05-31. It names one FDA 510(k) clearance: K971452. Product: AXIOM MULTIX MP - Inended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 8395431. Reason: Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Ongoing
Initiated
2024-05-31
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AXIOM MULTIX MP - Inended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 8395431

Reason for recall

Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire

Data sourced from openFDA. This site is unofficial and independent of the FDA.