Recall Z-2183-2014— ORTHO-CLINICAL DIAGNOSTICS

Class II · Terminated

Class II FDA medical device recall by ORTHO-CLINICAL DIAGNOSTICS, initiated 2014-07-16, terminated 2017-03-06. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.. Reason: Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values..

Recalling firm
ORTHO-CLINICAL DIAGNOSTICS
Classification
Class II
Status
Terminated
Initiated
2014-07-16
Terminated
2017-03-06
Firm location
Bridgend, N/A, United Kingdom

Product description

VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.

Reason for recall

Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

Data sourced from openFDA. This site is unofficial and independent of the FDA.