Recall Z-2180-2015— American Medical Systems Innovation Center - Silicon Valley

Class II · Terminated

Class II FDA medical device recall by American Medical Systems Innovation Center - Silicon Valley, initiated 2015-05-28, terminated 2016-01-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Tool, Stripper W/BLD, 44S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.. Reason: validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products.

Recalling firm
American Medical Systems Innovation Center - Silicon Valley
Classification
Class II
Status
Terminated
Initiated
2015-05-28
Terminated
2016-01-19
Firm location
San Jose, CA, United States

Product description

Tool, Stripper W/BLD, 44S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.

Reason for recall

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

Data sourced from openFDA. This site is unofficial and independent of the FDA.