Recall Z-2174-2013— Advanced Sterilization Products
Class II · Terminated
Class II FDA medical device recall by Advanced Sterilization Products, initiated 2013-07-25, terminated 2013-11-27. It names one FDA 510(k) clearance: K071385. Product: STERRAD 100NX Cassettes, Product Code: 10144 The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.. Reason: The recall of the STERRAD 100NX Cassettes was initiated because Advanced Sterilization Products (ASP) does not have adequate data to support the effectiveness of the packaging's leak indicator for the entire duration of the STERRAD 100 NX System cassette's 15 month shelf life..
- Recalling firm
- Advanced Sterilization Products
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-07-25
- Terminated
- 2013-11-27
- Firm location
- Irvine, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
STERRAD 100NX Cassettes, Product Code: 10144 The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Reason for recall
The recall of the STERRAD 100NX Cassettes was initiated because Advanced Sterilization Products (ASP) does not have adequate data to support the effectiveness of the packaging's leak indicator for the entire duration of the STERRAD 100 NX System cassette's 15 month shelf life.